Black Friday 2020 Tuttnauer Autoclave Repair: What Causes The Add Water Error Code

It looks as if it ought to be fairly straightforward to restore, proper? Your Tuttnauer autoclave is providing you with an error code that tells you so as to add water, in order that in all probability means all it’s essential to do is add some water. However while you have a look, you see the reservoir is already full. Now what?

Your autoclave is one in every of your most essential items of kit in your workplace and if one thing goes fallacious it is comprehensible that you’d panic. You are pondering of all the cash it is going to price to ship it out for restore, the expense of renting a substitute, and also you would possibly even have to shut your workplace for some time.

When the sensors point out that it ought to be a simple repair – like simply including some extra water – and the straightforward repair does not work, it may be irritating. However do not panic but. Autoclaves solely look intimidating. They’re actually quite simple items of kit. There’s often a quite simple answer and more often than not you are able to do it your self.

Test the mannequin quantity in your autoclave. Tuttnauer’s hottest fashions are their E Collection autoclaves, together with the EZ9 and EZ10.

If in case you have an E Collection mannequin, then take a peek contained in the water reservoir and you will see a float. This float is connected to a water degree change which acts similar to a light-weight change – while you flip it on, the autoclave comes on, and while you flip it off, the autoclave turns off.

When every little thing’s working correctly the water degree within the reservoir lifts the float and it turns the change to the ‘ON’ place, making {an electrical} connection which permits the facility to run to your autoclave. When the water degree is simply too low, the float disconnects with the change, primarily turning your autoclave ‘OFF’ till you refill the reservoir.

When the water degree is simply too low you will see that ‘Add Water’ indicator, or error message, mild up in your show and, as a security precaution, the unit won’t come on once more till you fill it with water. Which leads us again to the unique drawback.

In your case, the ‘Add Water’ message is flashing however the reservoir is full. Now what?

If the reservoir is full and the float is shifting freely, then your drawback might be a faulty water degree change. To make certain, you’ll be able to run a easy continuity take a look at by eradicating the wires and placing a probe out of your meter (or continuity tester) to every wire.

With the meter set to check continuity:

  • When the float is within the Up place the meter ought to present continuity.
  • When the float is within the Down place the meter ought to present an open circuit.

As soon as you’ve got verified that the issue is within the change you’ll be able to cease worrying. All it’s essential to do is exchange the change and if you should use a wrench you are able to do it your self. After you exchange the water degree change in your Tuttnauer autoclave remember to run a take a look at cycle to verify every little thing’s been repaired correctly and also you’re again in enterprise.